MGH head 1:MGH head 1

From Martinos Center Anthropomorphic Phantoms
Revision as of 13:33, 24 March 2015 by Guerin (talk | contribs)
Jump to navigation Jump to search

Citation

Giorgio phant.jpg

If you use any of the material below, please cite: Nadine N. Graedel, Jonathan R. Polimeni, Bastien Guerin, Borjan Gagoski and Lawrence L. Wald (2015). "An anatomically realistic temperature phantom for radiofrequency heating measurements". Magnetic Resonance in Medicine 73(1):442–450

Download files

Click here to download the CAD model.

Phantom description

This phantom is based on a 1mm isotropic segmented head model which was labeled as described in MRI-based anatomical model of the human head for specific absorption rate mapping.

The segmented head data was converted into a simplified surface model with two compartments: Brain and other soft tissues. Bone and air (both low epsilon, low conductivity) are printed together out of plastic material with similar electrical properties.

= Special considerations

The sealing flange uses an o-ring.

Phantom figure CT MR.png